- Model 306臭氧校准源(器)
详细信息
品牌:美国2B Tech 型号:Model 306臭氧校准源(器) 加工定制:否 分辨率:见具体描述 测量精度:见具体描述 气体流量:见具体描述 基本误差:见具体描述 尺寸:见具体描述 mm 重量:见具体描述 kg 电源:见具体描述
The 2B Technologies Model 306 Ozone Calibration Source™ is a portable source of ozone that allows you to calibrate any ozone monitor - not just those manufactured by 2B Tech. The instrument scrubs ozone from ambient air and produces any mixing ratio of ozone in the range 30 to1,000 ppbv. The desired ozone concentration is chosen from the easy-to-use menu using a rotary select switch. The total output flow rate is 3.0 L/min volumetric, and the ozone mixing ratio is controlled so as to be independent of ambient temperature, pressure and humidity. You can attach the Ozone Calibration Source (OCS™) output directly to the inlet of any ozone monitor; the excess flow is vented through an ozone scrubber internal to the OCS. Although the instrument comes calibrated against a NIST-traceable ozone standard, the menu allows the user to change the calibration parameters of the ozone output if desired. For example, if you maintain an ozone reference instrument in the lab, you can use the OCS as a transfer standard for calibrating ozone instruments in the field relative to your lab standard.
The Ozone Cal Source™ is housed in the same small, rugged instrument case as the Model 202 and Model 205 Ozone Monitors™.
Some applications of the Model 306 Ozone Calibration Source™ include:
- Calibration of ozone monitors in the field. This highly portable transfer standard is especially useful for maintaining the calibrations of networks of ozone monitors.
- Testing of materials such as rubber, plastics and paint for effects of ozone exposure.
- Studies of the effect of ozone on plants where a calibrated source of ozone is required.
The 2B Technologies Model 306 Ozone Calibration Source™ Example Data
The following are data obtained for the calibration of a Model 202 Ozone Monitor, which had previously been calibrated against our NIST-traceable standard instrument. New target ozone concentrations were entered into the Ozone Cal Source every 5 minutes. As can be seen from the data, the new ozone concentrations are reached within 3 data points (30 s). Concentrations were stepped up and down twice to produce 4 calibration plots. Target ozone mixing ratios were 0, 50, 100, 200, 400, 600, 800 and 1,000 ppbv. There is an extra measurement at 150 ppbv in the rise side of the first set of "stairsteps".
Fig.1. Measured ozone concentrations for "stair step" changes in Ozone Calibration Source Output.
The following figure shows the 4 calibration curves produced from the 4 sets of increasing and decreasing ozone concentrations. The data points and lines for the 4 least squares fits, which are plotted in different colors, overlap so well that it is difficult to distinguish them. The equations for the linear regressions show that the slopes of the 4 individual calibration curves agree to better than ±1% and the intercepts agree to ±2 ppbv. Correlation coefficients (R2) are all 0.9999 or 1.0000.
Fig. 2. Calibration curves established from data of Fig. 2.
The 2B Technologies Model 306 Ozone Calibration Source™ Specifications
Method of Ozone Production UV Photolysis of O2 at 185 nm Ozone Concentration Range 0 ppbv and 30 ppbv to 1,000 ppbv Precision and Accuracy of
OutputGreater of 2 ppbv or 1% of ozone concentration Response time for Change in
Ozone Output Concentration30 s to reach 95% of concentration change Flow Rate 3.0 Liter/min volumetric Power Requirements 12 V dc or 120/240 V ac, 18 watt Size 3.5 x 8.5 x 11 inches (9 x 21 x 29 cm) Weight 5.6 lb (2.6 kg) - Calibration of ozone monitors in the field. This highly portable transfer standard is especially useful for maintaining the calibrations of networks of ozone monitors.
-
供应商的其他相关信息
查看更多